Simini F, Santos D, Artigas J, Gigirey V, Dibarboure L and Francescoli L. Measurement of knee articulation laxity by videofluoroscopy image analysis: CINARTRO. Med Imaging Radiol. 2017; 5:4. http://dx.doi.org/10.7243/2054-1945-5-4
Franco Simini1*, Dario Santos1,2,3, José Artigas1, Verónica Gigirey4, Luis Dibarboure4 and Luis Francescoli5
*Correspondence: Franco Simini simini@fing.edu.uy
1. Núcleo de Ingeniería Biomédica, Facultades de Medicina e Ingeniería, Universidad de la República - Montevideo, Uruguay.
2. Departamento de Rehabilitación - Biomecánica Clínica, Hospital de Clínicas, Montevideo, Uruguay.
3. Escuela Universitaria de Tecnología Médica, Universidad de la República, Montevideo, Uruguay.
4. Departamento de Imagenología Clínica, Hospital de Clínicas, Montevideo Uruguay.
5. Clínica de Ortopedia y Traumatología, Facultad de Medicina, Montevideo Uruguay.
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Several methods are used to evaluate knee articulation during movement, such as the Tibio Femoral Contact Point migration or the Instantaneous Centre of Rotation displacement and even the Moment Arm variations during extension. Such evaluation is important in case of Anterior Cruciate Ligament reconstruction or Total Knee Replacement, but unfortunately, it is only available in research settings. We have devised a novel clinical methodology based on serial images of knee movement videofluoroscopy, which are interactively analysed to allow automatic determination of kinematics parameters. The method, called CINARTRO, compares values with the contralateral knee and produces a standard Clinical Document Architecture file ready for Electronic Medical Record systems. To correct the “pin cushion” effect of X ray intensifiers, a 10 mm apart lead spheres phantom was developed. To validate the interactive selection of anatomical points, cadaver limbs were marked with lead markers and compared with points marked on the blind image. Anterior Cruciate Ligament (injured and reconstructed) and contralateral Moment Arms averages of 53mm, 39mm and 43mm respectively were found. Similarly the Total Knee Replacement patient`s Moment Arm was reduced to 33mm. These findings justify the use of our methodology to help clinicians during rehabilitation follow-up.
Keywords: Knee Kinematics, Tibio Femoral Contact Point, Quadriceps Moment Arm, Videofluoroscopy, CINARTRO
The knee articulation (KA) is a very complex structure of the human body, with several functions such as ensuring static stability during full extension of the leg, shortening of the leg to allow movement and adaptation to ground irregularities during gait and running [1]. Until recently, clinical evaluation of movement is the result of subjective appraisal, highly dependent upon the observer. A few quantitative and therefore objective methods are available in the literature, all circumscribed within research settings. One method is the determination of the Tibio Femoral Contact Point (TFCP) and its migration over the tibial plateau during flexo/extension [2,3].
The knee is subject to continuous movements and exercises in everyday life. After a serious lesion such as the rupture of the Anterior Cruciate Ligament (ACL) the knee joint kinematics are altered [1]. The yearly incidence of injured ACL is one every 3000 persons, which implies an ample field of clinical work to be performed. In the economy, it accounts for over a billion dollars in surgical and rehabilitation costs only in the USA [4]. Total Knee Replacement (TKR) is an effective way to treat knee arthropathies, for which it is considered the definitive treatment for symptomatic end-stage osteoarthritis of the knee [5]. TKR is a successful treatment for most pain and dysfunction conditions in patients [6,7]. Postoperatively, functional instabilities may develop which require careful and evidence-based rehabilitation. One of the most common ways to check the state of a re-constructed knee is the Lachman test [2]. This evaluation depends greatly upon the perception and experience of the physician or rehab technician. An objective, quantitative measurement is needed and no such instrument exist to help the clinicians. To address the lack of objective methods available for routine clinical settings, we set ourselves the goal of designing a simple way with the simplest possible instrumentation and software. The determination of knee arthrokynematics could be used as an objective method to evaluate ACL reconstruction [3,4]. As an objective measurement, TFCP migration data, for instance, can help to put in place rehabilitation strategies and are suitable to monitor them, as well as to give feedback to the surgeon and Physical Rehabilitation Medicine staff. The methodology and the clinical instrument to be used for follow-up will be based on our prior medical device development experience [5,6] and the best available biomechanics research findings.
We have developed a way to enhance and supplement the subjective test with quantitative methods [7-9]. The aim of CINARTRO [7] (such is the name of our instrument, named as the contraction of “CINE” which stands for ”KINEMATICS” and ARTRO for “ARTHRO” meaning “joint”) is the collection of data and images to be processed into numerical results to help determine the condition of the patient after an ACL surgery [8]. The set of videofluoroscopy (VFC) images is the source of input data to the software, which interacts with the user to produce a relevant clinical report for the patient’s medical record. After preliminary evidence of knee dynamics evaluations [10], and developing a routine for VFC capture of the movement [11], we describe in the present paper the proof of concept of CINARTRO including calibration with phantom and cadaveric limb, as well as pave the way for the measurement of the degree of knee tightness or laxity.
Use cases of the instrument to be designed
Specifications of CINARTRO software
The Software we develop should be able to read DICOM images produced by the X-ray equipment, either the C-arm or other real time X-ray equipment. Based on this information, it must produce clinically relevant evidence according to the following specifications:
Building block of the instrument CINARTRO
To record a moving joint we chose a sequence of VFC images at a rate of 15 images per second during a two-second extension movement. This method has the advantage over conventional Roentgen moving film analysis (once known as RSA – Roentgen Stereo photogrammetry Analysis) in the much lower ionizing radiation delivered to the patient. The disadvantage of using VFC is the optical distortion inflicted to the image by the curvature of the image intensifier [3]. Using clinical VFC available for angiocardiographic or gastroscopic dynamic exploration, the overall dose delivered was 250 microGy (J/Kg) absorbed for every two seconds movement. To obtain VFC series we have used a variety of X-ray equipment provided they had serial imaging capacity. In Figure 1 we use a C arm and in Figure 2 a gastric dynamic imaging system. The images are recorded in DICOM format, with single images ready for further processing in Bit Map Picture (BMP). We designed an original software to process the series of images, which consists of an interactive application to allow the user to select fiducial points on the images. The CINARTRO software shows on the image the homothetic places coming from the previous one to help the operator by suggesting points to be either confirmed or modified. The eleven anatomical points prompted by the software are shown in the left column of Figure 4. The first five points include three points on the Femoral Condyle and the two Tibial Plateau extremities. The second group includes two pairs of long bone surface points to determine the flexion angle. Finally, the last group comprises the Patella Extremity and the Anterior Tuberosity Notch, to allow the software to define the quadriceps ligament direction, and thus the Moment Arm.
Figure 1 : C-arm used to obtain videofluoroscopic (VFC) images of a moving knee joint.
Figure 2 : Patient with Total Knee Replacement (TKR) during CINARTRO Videofluoroscopic procedure to evaluate dynamic functional knee parameter in open chain, i.e. under no load.
Image quality control to reduce pin cushion effect
In order to correct the pin cushion distortion effect, we have designed a plastic device [20-22] with lead spheres separated by 10 mm in a two dimensional regular pattern (Figure 3). We called the device “Phantom” and it was placed in the sagittal plane the patients would be located subsequently. Due to the relative intensifier/source geometry, the X-ray image taken of the Phantom pattern appears distorted. By matching the theoretical position of every lead sphere with their real position, our software [3,20] is able to correct the operatordefined fiducial points. We have put in practice the method published by Baltzopoulos [3] which minimizes a global error function with linear algebra methods. By doing so, we obtain a set of parameters for the transformation of the VFC point coordinates into “corrected” points. The same transformation derived from the Phantom image is later applied to all operator selected points on patient knee joint images. The image analysis is therefore free from major geometrical distortion.
Figure 3 : Phantom specially designed and built to obtain an image of the lead sphere pattern, distorted as a result of intensifier/source geometry (Pincushion effect).
Figure 4 : Screen of CINARTRO software.
The error in TFCP migration was found to be less than 2% of the Tibial Plateau dimension in case the pincushion effect was not corrected [23]. A TFCP measurement of 40% of Tibial Plateau extension, with an uncertainty of 2% can be made better with pincushion distortion correction. After correction, the error has no practical relevance in direct and derived measurements.
Validation of operator point marking
In order to verify that the operator marks the anatomical points correctly, we surgically prepared a cadaveric lower limb with lead spheres in the exact places of the Tibial Plateau extremities and along the Femoral Condyle. Similarly, we also put spheres at the lower end of the Patella and at the Tuberosity Notch.
In addition, the cadaveric limb was marked with steel rods to measure the flexion angle (Figure 4), as a check for the angle measured by CINARTRO software when processing the images.
Comparing cadaveric marked points with operator selection on blind images, the mean difference between series is 5% for the TFCP migration [24]. Similarly, the MA values obtained differ by a mean value of 3% [24], (Figure 5).
Figure 5 : Anatomical points marked by the user with CINARTRO on a Total Knee Replacement (TKR) image of patient of Figure 2.
CINARTRO to evaluate ACL reconstructed patients
We used CINARTRO to evaluate ACL reconstruction of six subjects, after securing Ethics Committee project approval. Table 1 shows the results in terms of percentage of the Tibial Plateau to normalize dimensions across patients. The MA is increased when the ACL is injured, returning to close to normal after reconstruction.
Table 1 : Quadriceps moment arm of 6 patients.
CINARTRO produces a clinical document, which includes a graphic representation of the TFCP migration and the MA as the extension progresses. Figure 6 is an example of such representation where the migration of the TFCP is shown to span from 42% to 60% of the Tibial Plateau length.
Figure 6 : Clinical results for the Patient Report.
CINARTRO to evaluate total knee replacement
We used CINARTRO to evaluate a Total Knee Replacement (TKR) patient (Figure 2), after obtaining his informed consent and Ethics Committee approval. Table 2 shows the results of the center of rotation proxy migration in terms of percentage of the Tibial Plateau compared to normal patients. The MA is also shown in Table 2 with the same normal reference values.
Table 2 : Total knee replacement moment arm.
When the ACL is repaired, the rehabilitation process aims at strengthening muscular action and neuromuscular control, in order to preserve as much as possible the original functions present prior to the lesion. In particular, the initial training phases are critical in strengthening quadriceps without damaging the reconstructed ACL [13]. When the ACL is injured, the quadriceps force is reduced [8]. We have shown here that the MA is slightly increased when ACL is injured (Table 1), which can be considered a compensation of the reduction in quadriceps strength, trying to keep the Moment constant (remember that Moment is the product of Force by MA).
The increase of MA secondary to ACL injury is shown in the same Table 1 to be reduced after ACL reconstruction, in an attempt to recover normal values. An average over the four values of healthy knee (across extension angles 135 º, 150 º, 165 º &180 º) is MA=43 mm, quite similar to established findings [25]. When injured, the mean MA=53 mm. After reconstruction, the surgeon contributes with a resulting mean MA=39 mm. This can be interpreted as a tighter knee joint to be taken over by rehabilitation. Quadriceps strength must be increased as a measure to counteract the surgical reduction of MA. A non-repaired ACL is probably compatible with a weakened quadriceps, but a repaired ACL (with lower MA) must be moved by a powerful quadriceps.
In case of the TKR patient, the MA (33 mm) is lower than the normal mean MA (from Table 1 mean MA=43 mm). As in the ACL reconstructed cases, the TKR patient also requires professional rehabilitation.
Clinical practice of ACL reconstruction follow-up has demanded for many years a practical instrument to benefit from research in joint kinematics [26]. On one side research explained the dynamics of the consequences of ACL rupture [12] and the partial reconstruction obtained with different surgical techniques. This explanation remained within the realm of academic work, while clinicians had no other way to record the evolution of a repaired ACL but static images and simple clinical procedure such as the KT1000 or Lachman test. By developing CINARTRO, we are suggested an objective procedure to record the result of ACL reconstructive surgery immediately after inflammation lowers, and later at intervals during physiotherapy and rehabilitation [8].
The X-ray system used for the present preliminary proof of concept of CINARTRO is available but cumbersome to use and expensive. We are working towards the design of a dedicated compact X-ray device, probably derived from veterinary equipment [20], which will allow us to design a simple, portable CINARTRO for routine clinical practice.
Future work will address the measurement of forces involved in the intact/torn/reconstructed ACL settings. By doing so, we will be able to further explore the hypothesis of quadriceps strength reduction with injured ACL or TKR. Consequently, rehabilitation will be asked to restore higher performance muscles to deal with reconstructed ACL and therefore with often shortened MA.
The authors declare that they have no competing interests.
Authors' contributions | FS | DS | JA | VG | LD | LF |
Research concept and design | √ | √ | -- | -- | -- | -- |
Collection and/or assembly of data | -- | √ | √ | -- | -- | -- |
Data analysis and interpretation | √ | √ | √ | √ | √ | √ |
Writing the article | √ | √ | √ | -- | -- | - |
Critical revision of the article | √ | √ | √ | √ | √ | √ |
Final approval of article | √ | √ | √ | √ | √ | √ |
Statistical analysis | √ | √ | -- | -- | -- | -- |
The authors thank Dr. Teresa Camarot, Director of the Rehabilitation and Physical Medicine Department. The authors also thank Dr. Eduardo Olivera for active cooperation from the Anatomy Department, and Prof. Vìctor Ezquerra of the Imaging Technical Department. The authors finally thank Ignacio Ferrer, Engineering Student as well as Andrés Rey for contributions in procedures and programming tasks. Mrs. Susana Martìnez is acknowledged for excellent administrative support.