Robertson M and Gregory R. Concurrent validation of the tandem walk test as a measure of dynamic walking balance in a healthy population. Phys Ther Rehabil. 2017; 4:12. http://dx.doi.org/10.7243/2055-2386-4-12
Marc Robertson1* and Robert Gregory1†
*Correspondence: Marc Robertson robertsonm1@southernct.edu
†The author contributed equally to this work.
1.Southern Connecticut State University, New Haven, Connecticut, USA.
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/3.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Background: The purpose of this study was to determine the possible concurrent validity of the tandem walk test as a measure of dynamic walking balance.
Methods: Forty-nine healthy participants, age range 20-75 years (mean age 40.1 ± 17.1 years), performed the ten step tandem walk test and two clinical tests of dynamic balance using the Biodex Balance System (Biodex Medical Systems, Shirley, NY) dynamic limits of stability protocol and measures of gait variability assessed through gait center of pressure, spatial, and temporal parameters derived from an instrumented treadmill (h/p/cosmos sports & medical gmbh, Nussdorf-Traunstein, Germany; Zebris FDM-T, Isny im Allgau, Germany).
Results: A moderate-high correlation was demonstrated between the ten step tandem walk test eyes open and measures of gait variability, step width (r=0.64 p<0.05) and step time variability ( r=0.67 p<0.05), age (r=0.69 p<0.05), and Biodex Balance System dynamic limits of stability (r=0.81 p<0.05). A moderate relationship was found between the ten step tandem walk test eyes open and measures of gait center of pressure variability (r=0.52 p<0.05). Correlation coefficient values between the ten step tandem walk test eyes open and Biodex Balance System dynamic limits of stability, gait temporal, center of pressure, and spatial parameters demonstrated increased strength when adjusted for age.
Conclusion: The ten step tandem walk test demonstrated a moderate to high relationship with clinical tests of dynamic balance: Biodex Balance System dynamic limits of stability and gait center of pressure, spatial, and time parameters in a healthy population demonstrating concurrent validity as an objective measure of dynamic walking balance in a healthy population between the ages of 20 and 75 years.
Keywords: Postural balance, validation studies, gait, tandem walk, tandem walk test
Dynamic walking balance is an elusive construct, yet to be quantitatively identified by a gold standard measurement [1]. Lark & Pasupuleti [2] referenced any situation that maximally narrowed an individual’s base of support (BOS) maximally challenged her/his dynamic balance, therefore,walking tests that serve to maximally narrow an individual’s BOS, theoretically, should maximally challenge her/his dynamic walking balance. Many laboratory tests and components of functional performance test batteries, such as the Fullerton Advanced Balance Scale [3,4], the LIFE pilot study protocol [5], Good Walkers Test [6], and the Established Populations for Epidemiologic Studies of the Elderly (EPIDOS) Prospective Study balance protocol [7] reported to test dynamic walking balance, most frequently used to predict those at risk for falls. These test batteries have been validated specifically for the elderly population which has been identified as a population predisposed to impaired dynamic walking balance and falls [8-10] hence, limiting these tests’ capacity towards accurately depicting dynamic walking balance in a healthy, younger population. Measures of gait variability (center of pressure (COP) [11-13], temporal [14-18], and spatial [19-22] measures) and balance systems, such as the Biodex Balance System (Biodex Medical Systems, Shirley, NY) have commonly been used in the laboratory setting in an attempt to quantify and qualify dynamic walking balance.
Variability of the COP trajectory upon walking has demonstrated the capacity to discriminate between individuals with stable and unstable gait and has been used as a measure of dynamic walking balance [11-13]. COP variability was sensitive to age related changes in gait patterns between middle aged and young female healthy participants [11], utilized as a composite measure of center of mass (COM) and COP to quantify balance control during gait in fall prone and healthy elderly walkers [12], and successfully used to characterize measures of postural control/dynamic balance under both normal and neurologically impaired gait [13].
Temporal gait parameters such as stride time, swing time, and step time variability demonstrated the ability to discriminate between fallers and non-fallers and were considered representative of dynamic walking balance [14,15]. Swing time variability, independent of gait speed, was found to be a valid measure of gait rhythmicity and walking balance [16]. Stride time variability was correlated (r=0.89 p<.0001) with swing time variability and suggested as an indicator of dynamic walking balance [15]. Low stride time variability equated to safe gait and increased gait stability [17]. Participants with neuropathic gait demonstrated increased step time variability which was associated (r=0.62 p=.032) with unstable walking [18].
Variability within spatial gait parameters, such as step width, step length, and stride length, has equated an increase in variability to an increased risk of falls in elderly individuals [19-22].
The Biodex Balance System (Biodex Medical Systems, Shirley, NY) (BBS) has been used to quantitatively characterize dynamic balance in a multitude of settings. The Biodex Balance System dynamic limits of stability test (BBS DLOS) was used to discriminate dynamic balance between individuals with rheumatoid arthritis and controls [23], very old male and female nursing home residents [24], and for the quantitative assessment of balance in individuals with spinocerebellar ataxia [25]. Clinical utility of the BBS DLOS was determined through intra-tester, inter-tester [26] and test-retest reliability [27]. The BBS DLOS was compared to the NeuroCom Smart Balance Master system (NeuroCom International, Clackamas, OR) to assess the dynamic balance discrimination capacity between the two systems [1].
Accurate quantification of gait stability requires sophisticated instrumentation not available to most clinicians [2]. Laboratory measures of dynamic walking balance demonstrate limited clinical utility due to prohibitive cost and portability limitations. The tandem walk test (TWT) has been historically utilized as a reliable measure of dynamic walking balance since 1938 by Dain who measured dynamic balance in boys between the ages of 13 and 18 via a beam walking test and later utilized and refined by Graybiel & Fregly in 1965 [28-30].
The TWT has frequently been included in standard neurological examinations to indicate neuro-motor competence and dynamic balance capacity [31-33]. The TWT test has been utilized as the dynamic walking balance component of several validated test batteries regarding fall prediction in the elderly, such as the Fullerton Advanced Balance Scale [3,4], the LIFE pilot study protocol [5], Good Walkers Test [6], and the EPIDOS Prospective Study balance protocol [7]. The TWT has been utilized as a test of dynamic walking balance to determine gait stability and pathological quantification including prediction of falls in the elderly population [3,9,34-40], diagnose ataxia and other vestibular disorders [29,41], test for vertigo [42], tropical ataxic neuropathy [43], traumatic brain injury [40], peripheral neuropathy [44], and functional mobility [45-48]. The TWT is considered a clinically efficacious test of dynamic balance in medical text books [29,49] and for research purposes [2,50]. The TWT has been utilized as a measure of dynamic walking balance in both clinical and research settings.
The TWT is a simple, commonly used clinical assessment for the purpose of testing dynamic walking balance in patients and study participants. To date, no studies have served to validate the TWT as a clinically efficacious measure of dynamic walking balance in the general population [44], yet there has been a strong assumption that the TWT is a valid measure of dynamic walking balance due to its propensity to maximally narrow individuals’ walking BOS [2].
Therefore, the purpose of this investigation was to determine the concurrent validity of the ten step TWT eyes open (TSTWT EO) [8] as a tool for determining dynamic walking balance in a cross-sectional healthy population between the ages of 20 and 75 years to support the hypothesis that the TSTWT EO demonstrates concurrent validity as a test of dynamic walking balance due to its association with laboratory-based tests of dynamic balance. Concurrent validation of the TSTWT EO was tested through examination of its relationship with several laboratory based measures of balance, including the modified Fall Risk Questionnaire (mFRQ) [51], BBS DLOS, and COP, temporal, and spatial measures of gait variability quantified by a h/p/cosmos (h/p/cosmos sports & medical gmbh, Nussdorf-Traunstein, Germany; Zebris FDM-T, Isny im Allgau, Germany) instrumented treadmill with a Zebris pressure platform utilizing Noraxon (Noraxon U.S.A. Inc., Scottsdale, Arizona) myopressure software.
Participants
This cross-sectional study was conducted at Southern Connecticut State University (SCSU) in greater New Haven County of Connecticut, USA and was approved by the Southern Connecticut State University Institutional Review Board (IRB #15-139). Forty-nine healthy participants from New Haven County, Connecticut (USA) between the ages of 20 and 75 years were recruited for voluntary test participation. All participants completed written informed consent prior to study participation.
Study participant inclusion criteria were: Participants had to be between the ages of 20-75 years old, walk without an assistive device (cane, walker, crutches), no lower extremity or trunk surgeries within the past year, no acute illness or traumatic injury, no unstable health conditions, and had to be able to walk for greater than or equal to four minutes continuously without help/assistance.
Procedures
Tandem walk test
A fifteen foot long, one inch wide piece of white training tape was placed along a level concrete floor approximately two feet away from and parallel with a wall. The participant was asked to walk forward along the taped line, placing one foot directly in front of the other, as per the heel of one foot directly in front and in contact with the toes of the contralateral foot. Each participant attempted two trials, two sets each for TSTWT EO then TSTWT eyes closed (EC) per trial. Each participant was asked to tandem walk a maximum of ten steps at a comfortable pace with his/her arms folded across his/ her chest without performing a test error including forward progression stoppage, sidestepping, opening the eyes (if applicable), unfolding arms, not maintaining a parallel foot position with the taped line (when applicable), and/or requiring gait belt assistance to maintain an upright position [29]. Test protocol was demonstrated prior to the participant’s attempts. Maximum number of steps achieved before error was recorded.
Gait data (COP, spatial, temporal measures) and gait phase parameters
COP, in relation to dynamic walking balance, is the focal point of location of the vertical ground reaction force vector relative to thefoot in contact with the supporting surface which elicits a COP trajectory as the foot translates on the supporting surface [11,13]. The degree of variability within the COP trajectory has been used to quantify static and dynamic postural control in both normal and pathological populations with increased unilateral and side to side variability of the COP trajectory indicative of reduced dynamic walking balance capacity [11-13]. COP measures [ length of gait line left/right (mm) which described the trajectory of the pressure center, accounted for all trajectory planar deviations, and assessed for individual ground contacts on one side of the body; single support line left/right(mm)quantified as the average length of the unilateral trajectory line which described the trajectory of the pressure center and accounted for all ground contacts; and single support line percent difference which described the single support line side to side differences] collected at 300Hz, gait spatial measures (cm) [step length coefficient of variation(CV) left/right; step length percent difference; step width; and step width CV], temporal measures (s) [step time CV; stride time CV; and swing time CV], and gait phase parameters [double stance percentage, cadence, and gait speed] were obtained from participants while walking on a h/p/cosmos-Zebris Medical instrumented treadmill/force platform (h/p/cosmos sports & medical gmbh, Nussdorf- Traunstein, Germany; Zebris FDM-T, Isny im Allgau, Germany) with Noraxon Myopressure analytical software (Noraxon U.S.A. Inc., Scottsdale, Arizona). Participants were asked to walk at a comfortable, self-selected speed for four minutes with data, described above, collected over the last 60 seconds of the trial. Each participant was asked not to talk and to look straight ahead while walking on the treadmill. Treadmill speed was determined by advancing the treadmill speed one kilometer per hour until a self-reported comfortable speed was determined; the four minute protocol was then initiated. The participant walked at his/her self-selected pace for three minutes to allow for treadmill walking acclimation with an additional one minute data collection period. The participant was assisted off of the instrumented treadmill and escorted to a chair for three minutes of rest.
Biodex Balance System (BBS) Dynamic Limits of Stability (DLOS) (Biodex Medical Systems, Shirley, NY)
Participants were provided a verbal explanation and physical demonstration of the dynamic limits of stability (DLOS) test protocol. Participants were asked to step on the BBS stance platform for support rail and biofeedback display adjustment to ensure safety and comfort. Participant foot placement was adjusted on the BBS platform as standardized by BBS normative criteria based on participant height. One practice set, which consisted of three untimed practice trials, was performed prior to data collection trials. Participants stood on an unstable platform, shifted, and controlled their COG within their moving BOS in an attempt to direct a cursor on a biofeedback display from a centralized target to eight perimeter targets as quickly and with as little path deviation as possible. DLOS data were gathered over the course of three sets, each set consisted of three untimed trials with a three minute rest interval between each set (52). Balance platform stability was set at level eight of a 12 point scale with 12 being most stable and one being least stable. An overall balance index was used for data analyses.
Statistical analysis
Pearson’s product-moment correlation coefficient (r) was calculated to describe the relationship between the TSTWT EO and the BBS DLOS overall index, measures of gait variance (COP, temporal, and spatial measures) primarily expressed as coefficient of variation (CV)(standard deviation/mean*100), primarily used to standardize unilateral data (with the exception of step width CV) and account for associated variability in the referenced data set, and percent difference, used to assess side to side symmetry, TSTWT EC, gait phase parameters (gait speed, cadence, double stance), and participant BMI and age. Linearity of data was determined from scatter diagram analysis. Multiple linear regression analyses were calculated to adjust for participant age. All analyses were performed using SPSS (version 21). Statistical significance was determined at an alpha value of less than 0.05 (α<0.05).
Forty-nine volunteers successfully completed the current test protocol. See Table 1 for participant characteristics.
Table 1 : Participant Characteristics.
Pearson’s product-moment correlation coefficient (r) was performed between the TSTWT EO and research variables including the TSTWT EC, BBS DLOS Index, and the mFRQ demonstrating p< 0.05 with a high positive correlation [53] demonstrated with the TSTWT EC (r=0.78), BBS DLOS overall index (r=0.81), and a moderate positive correlation [53] with the mFRQ (r=0.61) (Table 2).
Table 2 : Pearson’s Product Moment Correlation Coefficients between TSTWT EO, BBS DLOS, & TSTWT EC and Measures of Dynamic Walking Balance.
A moderate positive [53] relationship was demonstrated between the TSTWT EO and spatial measures of gait (cm); step width percent difference (r=0.61), step width CV (r=0.64), and step length CV left and right (r=0.48 and r=0.49, respectively). A moderate positive [53] relationship was demonstrated between the TSTWT EO and temporal measures of gait (s); step time CV (r=0.67) and swing time CV (r=0.57) (Table 2).
A low positive [53] to moderate positive [53 ]relationship was demonstrated between the TSTWT EO and COP measures (mm); length of gait line left and right (r=0.35 and 0.31, respectively), length of gait line CV (r=0.52), single support line left and right (r=0.51 and r=0.52, respectively), single support line CV (r=0.48), and single support line percent difference (r=0.31 p=0.05). A moderate positive [53] relationship was demonstrated between the TSTWT EO and participant BMI (r=0.50) and a moderate-high positive [53] relationship was demonstrated between the TSTWT EO and participant age (r=0.69) (Table 2).
Multiple linear regression analyses were performed to adjust data for participant age. A high positive [53] relationship was demonstrated between the TSTWT EO and the BBS DLOS, TSTWT EC, step width CV, and step width percent difference (r=0.86, r=0.84, r=0.81, and r=0.75, respectively). A high positive [53] relationship was demonstrated between the TSTWT EO and the mFRQ (r=0.74), COP single support line left and right (r=0.73), COP length of gait line right (r=0.73), double stance percentage (r=0.72), step time variability (r=0.71), single support line difference (r=0.71), gait cadence (r=0.71), step length difference (r=0.70), and COP length of gait line left (r=0.70) (Table 3).
Table 3 : Pearson’s Product Moment Correlation Coefficients Adjusted for Age.
The TWT has historically been utilized as a reliable measure of dynamic walking balance in both clinical and research settings, yet to date, no studies have served to validate the TWT as a clinically efficacious measure of dynamic walking balance in a healthy population. The purpose of this investigation was to determine the concurrent validity of the ten step TWT eyes open (TSTWT EO) [8] as a tool for the assessment of dynamic walking balance in a cross-sectional healthy population between the ages of 20 and 75 years. The results from the current study supported the hypothesis that the TSTWT EO demonstrated concurrent validity as a test of dynamic walking balance in a cross-sectional, healthy population. The walk on floor eyes closed (WOFEC) version of the TWT had been validated, validity coefficient of .83, as a discriminatory test of walking dynamic balance in its ability to identify participants with labyrinthine defective ataxia [29], as a component of balance test batteries primarily utilized for elderly populations for the purpose of fall prediction [3-7], but had not been validated as an individualized measure of walking dynamic balance in a healthy population within the age range of 20 and 75 years. The TSTWT EO demonstrated moderate to high positive [53] correlation values with commonly used laboratory measures of dynamic balance (BBS DLOS, gait COP, time, and spatial measures, and the mFRQ) in a healthy population between the ages of 20-75 years, with increased strength of association demonstrated when values were adjusted for participant age.
COP parameters
COP trajectory has been used to measure static and dynamic postural control and as a balance index for normal and pathological populations [11,13]. Bizovska et al. [11] demonstrated a difference in COP trajectory upon walking between healthy young and middle-aged women, alluding to age-associated reduced dynamic walking balance and the efficacy of gait COP trajectory parameters in classifying dynamic walking balance. Lee & Chow [12] investigated COM and COP inclination angles, derived from both sagittal and frontal plane trajectories, in elderly walkers with a history of falls or unstable gait. Increased COM-COP frontal plane trajectory excursion identified elderly participants with dynamic walking balance abnormalities [12]. COP measures reviewed in the present study were single support COP trajectory line, left and right, and length of COP trajectory gait line, left and right. A moderate correlation was demonstrated between theTSTWT EO and single support COP trajectory line, left and right (r=0.51 p=0.001 and r=0.52 p=0.0007, respectively). The strength of relationship increased to a high-positive [53] correlation when adjusted for age (r = 0.73 p=0.03 left and right). A low-positive [53] correlation was determined between the TSTWT EO and length of COP trajectory gait line, left and right (r=0.35 p=0.03 and r=0.31 p=0.05, respectively) yet the correlational strength increased to a high-positive [53] relationship when data were adjusted for age (r=0.70 p=0.02 left and r=0.73 p=0.006 right). This low to high positive relationship between COP trajectory measures and the TSTWT EO implied that the relationship between dynamic walking balance and the TSTWT EO was strong when age was accounted for and low to moderate when age was not considered.
Temporal parameters
Irregular temporal parameters of gait have been associated with aging, falls, and a variety of diseases [18]. Step time variability has been associated gait pattern changes identified with aging, traumatic brain injury, diabetes mellitus associated peripheral neuropathy, Parkinson’s disease, Alzheimer’s disease, affective psychiatric disorder, fall risk, and falls [18]. Richardson et al. [18] studied the relationship between step length and step time variability on both smooth and irregular walking surfaces in older participants with neuropathy. An inverse relationship was reported between step length and step time variability, which indicated that as step length decreased, as found with reduced gait stabilization, step time variability increased linking increased step time variability with reduced walking stability, hence, reduced walking balance [18]. The TSTWT EO demonstrated moderate [53] (r=0.67 p=0.007) and high-positive [53] (r=0.71 p=0.000) correlations with step time CV relative to unadjusted and data adjusted for participant age, respectively. Frenkel-Toledo et al. [16] found an independent relationship between gait stability and swing-time variability, hypothesized as a valid indicator of gait steadiness and fall risk, indicative of dynamic walking balance. A moderate-high positive[53] relationship, without and with age adjustment (r=0.57 p=0.02, r=0.71 p=0.000, respectively) was found between swing-time CV and the TSTWT EO, indicating a moderate to high positive [53] relationship between dynamic walking balance and the TSTWT EO. Gait stride-to-stride fluctuations have been associated with falls, individuals at risk for falls, and normal aging [14,15]. Community- dwelling older adults demonstrated a fivefold increase in the possibility of falls upon only a moderate increase in stride time variability, indicative of gait instability and walking dynamic balance [14,15]. The TSTWT EO demonstrated a moderate [53] (r=0.43 p=0.04) correlation with stride time CV and a high positive [53] (r=0.70 p=0.000) correlation when adjusted for participant age, demonstrating a strong association between the TSTWT EO and dynamic walking balance.
Spatial parameters
Stride to stride variability has been associated with risk of disability [19] and falls [20]. Brach et al. [19,20] found that both excessive and reduced step width variability were indicative of gait instability as the ability to modify step width with gait was shown as a compensatory mechanism used to stabilize the COM under varying gait conditions, such as altered speed and/or uneven terrain. An age-associated increase in step width variability allowed for accurate participant age-range identification, which associated step width variability with risk for future falls [15,22]. The TSTWT EO demonstrated a high positive [53] correlation with step width with and without age adjustment (r=0.75 p=0.01 and r=0.61 p=0.006, respectively). A high positive correlation existed between the TSTWT EO and step width CV when data was adjusted for age (r=0.81 p=0.000), which suggested a strong association between the TSTWT EO and dynamic walking balance.
Biodex Balance System Dynamic Limits of Stability
The TSTWT EO demonstrated a high positive [53] correlation with the BBS DLOS test both prior to and with age-adjusted participant data ( r=0.81p=0.000 and r=0.86 p=0.000, respectively). Aydog et al. [23] compared dynamic balance between two groups of participants, healthy controls and participants with lower extremity involved rheumatoid arthritis (RA) and found dynamic balance was negatively affected in individuals with RA compared to healthy control participants which was objectively demonstrated using the BBS DLOS [23]. Mohan et al. utilized the BBS DLOS system to quantify dynamic balance in individuals with spinocerebellar ataxia in order to determine the efficacy of rehabilitative interventions [25]. Dynamic balance was compared between participants with diagnosed spinocerebellar ataxia and age matched healthy controls [25]. The BBS DLOS was reported to be the best predictor of pathological dynamic balance [25]. Pickerill & Harter attempted to establish the concurrent validity of two DLOS balance systems, BBS and the NeuroCom Smart Balance Master (Clackamas, OR) and found a low positive, yet non-significant correlation (r=0.33 p=0.12) between the two systems for overall directional balance control measures [1]. The non-significant low correlation was attributed to differences in testing protocols [1]. The BBS DLOS provided a greater challenge to participant’s functional stability due to its unstable standing base and was reported to uniquely challenge dynamic balance, yet Pickerill and Harter [1] found the NeuroCom (Clackamas, OR) to be more reliable as a LOS measure. Pickerill & Harter [1] reported the NeuroCom LOS maintained a fixed BOS, the BBS DLOS maintained foot position, yet the unstable stance platform of the BBS DLOS system continuously altered participant’s BOS resulting in a greater dynamic balance challenge. Hinman [27] found the BBS to be a reliable instrument for use in testing and quantifying dynamic balance in a healthy population. Limits of stability testing using the BBS was more difficult for participants than LOS testing using the NeuroCom due to the BBS’s unstable stance platform [27]. Hinman [27] suggested that dynamic balance testing using the BBS LOS test may have been too difficult for a participant population with pre-existing balance impairments as a percentage of test participants were unable to complete the DLOS tests without using upper extremity assistance to prevent complete loss of balance. Schmitz & Arnold [26] determined the intertester and intratester reliability values for the BBS DLOS for dynamic balance through a unilateral stance protocol based on anterior to posterior (A-P), medial lateral (M-L), and overall stability indices. Intertester and intra-tester ICC stability scores ranged from .70 to .42 and .80 to .43, respectively [26]. Schmitz & Arnold [26] concluded that the BBS was a reliable tool for testing dynamic stability. The BBS DLOS system appeared to accurately and uniquely depict dynamic balance in both pathological and healthy populations. The strong association found between the TSTWT EO and BBS DLOS suggested a strong relationship between dynamic balance and the TSTWT EO. Despite methodological controls built into this research design, a degree of subjectivity in regards to TSTWT LOB assignment may have influenced test results. Not with standing the reported clinical efficacy of the TSTWT EO as reported in the current study, careful consideration regarding clinical application of this study’s findings should be applied to dynamic walking balance test selection as the participant population was limited to a healthy population and did not include a clinical population. The propensity of the TSTWT EO to maximally narrow an individual’s BOS may overly challenge a clinical population with diminished dynamic walking capacity. The purpose of this study was to validate the TSTWT’s capacity to assess dynamic walking balance in a healthy population, not a clinically balance impaired population. A future direction for this research should entail usage of similar test methodology in a participant population with known diminished dynamic walking capacity.
There appeared to be a moderate to high positive [53] relationship between theTSTWT EO and commonly used laboratory measures of dynamic balance (BBS DLOS and gait COP, temporal, and spatial parameters) in a healthy population between the ages of 20-75 years, with increased strength of association demonstrated when values were adjusted for participant age. The TSTWT EO did not demonstrate criterion validity as a measure of walking dynamic balance since no criterion standard regarding dynamic walking balance has been identified, yet, due to the strength of association between the TSTWT EO and commonly used measures of dynamic walking balance, the TSTWT EO demonstrated convergent-construct validity as a measure of walking dynamic balance. The TWT has historically been utilized as a measure of dynamic walking balance due to its propensity to maximally narrow the base of support of the tested individual [2], but theTSTWT EO was never validated as a test of dynamic walking balance in a healthy population between the ages of 20 and 75 years. The TSTWT EO was found to be a clinically efficacious quantitative and qualitative measure of dynamic walking balance which demonstrated significant clinical utility compared with the afore-mentioned laboratory tests utilized as tests of dynamic walking balance considering cost and portability.
BBS: Biodex Balance System
BMI: Body mass index
BOS: Base of support
cm: centimeters
COM: Center of mass
COP: Center of pressure
CV: Coefficient of variance
DLOS: Dynamic limits of stability
LOS: Limits of stability
mFRQ: Modified Fall Risk Questionnaire
mm: millimeters
s: seconds
TWT: Tandem walk test
TSTWT EC: Ten step tandem walk test eyes closed
TSTWT EO: Ten step tandem walk test eyes open
USA: United States of America
The authors declare that they have no competing interests.
Authors' contributions | MR | RG |
Research concept and design | √ | √ |
Collection and/or assembly of data | √ | -- |
Data analysis and interpretation | √ | -- |
Writing the article | √ | -- |
Critical revision of the article | √ | -- |
Final approval of article | √ | √ |
Statistical analysis | √ | -- |
The authors would like to thank Dr. Robert Axtell for his advisory and editorial contribution to this project.
Editor: Catherine Ortega, University of Texas Health Science Center, USA.
Received: 29-Jun-2017 Final Revised: 01-Sep-2017
Accepted: 30-Sep-2017 Published: 15-Oct-2017
Robertson M and Gregory R. Concurrent validation of the tandem walk test as a measure of dynamic walking balance in a healthy population. Phys Ther Rehabil. 2017; 4:12. http://dx.doi.org/10.7243/2055-2386-4-12
Copyright © 2015 Herbert Publications Limited. All rights reserved.